Whole Blood Separation

D-23® Whole Blood Separation Media

Row filter-holdings-swoopty Shape Decorative svg added to bottom <

MATERIAL CLASS:
Whole blood separation - Coated agglutinating chemistry

Grade LF-D-23®
Grade CF-D-23®

Coated Whole Blood Separation Media.

Now Available in 2 thicknesses to improve device optimization

Grade D-23® is a high rate, high plasma yield blood separator for both lateral flow and vertical flow immunoassays. This media performs efficiently at a broad range of whole blood sample volumes, including <100µL which further expands your device design capabilities.

Material originates as our D-23® base glass borosilicate glass microfiber filter media containing a proprietary acrylic binder system. The base glass material, although an effective whole blood separator alone, is then treated with a novel coating technology which yields high quality plasma with surprising efficacy, speed and with little-to-no red cell hemolysis. Now available in 2 thicknesses for further optimization of your device design.

Plasma yield of >40%, at a rate of 30-120 seconds is achieved via the agglutinating chemistry and coating technology developed by a CRO in the United States whose specialty is blood separation and device development. This market proven media is currently included as a component in successfully commercialized and patented devices with 510K filings.

The agglutinating coating can be altered to best fit the application, separation rate and whole blood sample size for your device design.

Shown here are 4 titrations of the agglutinating chemistry (using no buffer) illustrating the plasma yield for similar Hct samples by volume and time in a simple lateral flow design.

Plasma yield zone
Resultat plasma yield zone
Blood

Material demonstrates high uniformity in formation and density allowing reliable migration rates, in both cross and lateral flow applications. With little caliper deviation the media is ideal for challenging lamination applications and micro fluidic devices.

Uniform density and stability is ideal for automated device assembly. This media demonstrates excellent lot-to-lot reproducibility - CofC and CofA provided with shipments – Manufactured in an FDA approved facility and available as rolls, reels, sheets and semi-finished converted configurations in both pre and post coated versions.

This media is also available as a base glass microfiber which performs basic whole blood separation via precisely engineered porosity and fiber matrix. In the natural uncoated configuration both caliper versions intrinsically demonstrate plasma yield of approximately 15% by volume at nominal Hct range. It is suitable for both lateral and vertical flow configurations and is highly consistent in density and caliper for application of your own micro-coating and dip/dry saturation chemistry.

3 plasma separation modes in 2 material thicknesses are now available containing proprietary agglutinating chemistry coating. In this coated versions demonstrate plasma yield >40%.

  • Coated for high yield agglutinating process, LATERAL FLOW
    Hct RANGE: NOMINAL = x1
    Hct RANGE: HIGH = x2
  • Coated for high yield agglutinating process, CROSS FLOW (vertical)
    Hct Range: NOMINAL = x1

Basic material physical characteristics:

Basis Weight
(g/m2)
Caliper Thickness
(mm)
Micron Retention
(µm)
Tensile strength
(lb)
Basic migration rate
D-23®
94.40.503-58.0 MD, 4.0 CD35 sec. / 4cm
D23®-TC1
70.00.3753.67.0 MD, 5.5 CD28 sec. / 4cm

Known competitive grades*:

  • GE Healthcare Whatman Fusion 5
  • GE Healthcare Whatman GF/DVA, LF1, MF & VF2
  • Ahlstrom / Pall Cytosep®
sample kit for coated D-23® variants

Please request your device developer’s sample kit for coated D-23® variants along with additional technical information.

Coated with high yield agglutinating process chemistry,
LATERAL FLOW

  • LF-D23-X1 (thick version, designed for nominal Hct range, lateral flow)
  • LF-D23-X2 (thick version, designed for high Hct range, lateral flow)
  • LF-D23-TC1-X1 (thin version, designed for nominal Hct range, lateral flow)
  • LF-D23-TC1-X2 (thin version, designed for high Hct range, lateral flow)

 

Coated with high yield agglutinating process chemistry,
CROSS FLOW

  • CF-D23-X1 (thick version, designed for nominal Hct range, cross flow vertical applications)
  • CF-D23-TC1-X1 (thin version, designed for nominal Hct range, cross flow vertical applications)